BABY INCUBATOR
Manufacturer:
SPACETECH HEALTHCARE LIMITED
Model: INFANTILE 800
The Infantile 800 baby incubator uses front and rear double-walls which allows heated air flow between the walls to conserve heat. Automatic air curtain design, when the access panel is opened, heated air flow is accelerated to prevent heat loss.
Trend for 2, 4, 8, 12, 24 hours and up to 7 days for air and skin temperature, humidity, oxygen and weight.
The Infantile 800 baby incubator can be operated from both sides of the cabinet. Special design allows closer and comfortable position to reach the newborn.
Key Features
-
8.4" LCD display relevant information can be observed easily.
-
One-liter water reservoir enables the to keep running for 24 hours with 85% humidity.
-
The cabinet is easy to remove and install. Special design makes every corner of the cabinet could be cleaned and sterilized completely.
-
All the operation can be completed at front side and don't need to move to other side.
-
Dual-CPU design to avoid single failure. Distinguish data collection and control from two CPU to make the Infantile 800 more reliable.
-
PPSU material is applied on water reservoir, so the water reservoir is able to be sterilized on 134℃ high temperatures.
GENERAL SPECIFICATION
Eight Tube accessing port.
18cm x 13cm access door size
45.8cm mattress to hood size
Mattress tilt +12 degree to -12 degree
Noise level: 45 dBa or below.
Battery backup: 30 minutes for support air circuit monitoring and alarms
TEMPERATURE CONTROL MODES
Skin temperature and air temperature control mode.
Skin control temperature range: 34°C – 37°C
Skin control override temperature range: 37.1°C – 38°C
Skin temperature display accuracy: ±0.3°C
Dual skin temperature monitoring (T1, T2)
Air temperature control range: 20°C – 37°C
Air temperature override control range: 37.1°C – 39°C
Air temperature display accuracy: ±0.3°C
Heater output: 0 – 100 %: (indicated in 10 levels)
Trend time: 24 hours of air temperature, skin temperature, relative humidity, oxygen concentration and heater power.
HUMIDITY
Humidification method: Boiling method
Humidity control mode: Servo controlled
Humidity control range: 30% to 95% Rh in 1% increment
Humidity operating control time: 24 hour max. 85%; 36°C.
Humidity control reservoir capacity: 1.5 liter (1500 ml)
Humidity display accuracy: ± 5% Rh.
O2 CONTROLLER
Oxygen control mode: Servo control
Oxygen control range: 21% to 65% in 1% increment
Oxygen display accuracy of full scale: ±2%
Oxygen control accuracy 100% calibration: ±3%
Oxygen control accuracy 21% calibration: ±5%
Oxygen display resolution: 1%
ALARM SAFETY SIGNALS
Visual and audible battery buffered/powered alarms for high and low skin & air temperature, skin probe disconnect, cutout, electronic circuit failure, fan failure, system failure, power failure, stuck key, sensor disconnect, sensor module failure, air flow probe failed, battery disconnect, sensor out of position, access panel open, oxygen cell difference, add water, humidity heater failure.
STANDARD ACCESSORIES
Skin temperature probe
IV pole
Replaceable bacterial filter
Mattress
OEM Warranty Certified
Turnkey Installation
24/7 Biomedical Support